Skip to search boxSkip to navigationSkip to main content

Repeatability of ocular biometry with IOLMaster 700 in subjects with clear lens

  • Fundación Destellos de Luz IBP
    ,
  • Instituto Mexicano del Seguro Social
Research Output:
Contribution to journal
Article
Peer-review

Publication metrics

Metrics

SciVal
FWCI
1.46
SciVal
Author count
3
SciVal
Citations
34
SciVal
Paper percentile
80
Scopus
Citations

Abstract

Purpose: The purpose of the study was to assess the reproducibility of ocular biometry using the IOLMaster-700 in a healthy population. Methods: This is a prospective, cross-sectional, observational reproducibility study. Ocular biometry was performed three times on each of 45 studied eyes. Flattest meridian (Kf) and the steepest meridian (Ks), central corneal thickness, axial length, anterior chamber depth, aqueous depth, lens thickness, and white-to-white distances were recorded. Reproducibility was evaluated using the coefficient of variation (CV), the within subject standard deviation, and the intraclass correlation coefficient (ICC). Results: There was a high reproducibility in all parameters; CV was between 0.3 and 1 %, and the ICC was higher than 0.87 in all measurements. Conclusions: IOLMaster-700 showed high reproducibility for ocular biometry.

Publication Information

Output type

Research Output:
Contribution to journal
Article
Peer-review

Original language

English

Pages from-to (Number of pages)

Pages 1133-1138 (6 pages)

Journal (Volume, Issue Number)

International Ophthalmology (Volume 37, Issue 5)

Publication milestones

  • Published - 01/10/2017

Publication status

Published - 01/10/2017

ISSN

0165-5701

Publication IDs

  • Scopus: 84992017544
  • WOS: 000412662600007