Repeatability of ocular biometry with IOLMaster 700 in subjects with clear lens
- ,
- Hugo A. Fuentes-de la Fuente,
- Ana V. García-Treviño
- Fundación Destellos de Luz IBP,
- Instituto Mexicano del Seguro Social
Publication metrics
Metrics
Abstract
Purpose: The purpose of the study was to assess the reproducibility of ocular biometry using the IOLMaster-700 in a healthy population. Methods: This is a prospective, cross-sectional, observational reproducibility study. Ocular biometry was performed three times on each of 45 studied eyes. Flattest meridian (Kf) and the steepest meridian (Ks), central corneal thickness, axial length, anterior chamber depth, aqueous depth, lens thickness, and white-to-white distances were recorded. Reproducibility was evaluated using the coefficient of variation (CV), the within subject standard deviation, and the intraclass correlation coefficient (ICC). Results: There was a high reproducibility in all parameters; CV was between 0.3 and 1 %, and the ICC was higher than 0.87 in all measurements. Conclusions: IOLMaster-700 showed high reproducibility for ocular biometry.
Publication Information
Output type
Original language
EnglishPages from-to (Number of pages)
Pages 1133-1138 (6 pages)Journal (Volume, Issue Number)
International Ophthalmology (Volume 37, Issue 5)Publication milestones
- Published - 01/10/2017
Publication status
ISSN
0165-5701Publication IDs
- Scopus: 84992017544
- WOS: 000412662600007