Comparison of biometric measurements obtained by the Verion Image-Guided System versus the auto-refracto-keratometer

  • Cecilio Velasco-Barona
    ,
  • Guadalupe Cervantes-Coste
    ,
  • Erick Mendoza-Schuster
    ,
  • Claudia Corredor-Ortega
    ,
  • Nadia L. Casillas-Chavarín
    ,
  • Alejandro Silva-Moreno
  • Asociacion para Evitar la Ceguera en Mexico
Research Output:
Contribution to journal
Article
Peer-review

Publication metrics

Metrics

SciVal
FWCI
0.58
SciVal
Author count
8
SciVal
Citations
5
SciVal
Paper percentile
57
Scopus
Citations

Abstract

Purpose: To compare the biometric measurements obtained from the Verion Image-Guided System to those obtained by auto-refracto-keratometer in normal eyes. Methods: This is a prospective, observational, comparative study conducted at the Asociación para Evitar la Ceguera en México I.A.P., Mexico. Three sets of keratometry measurements were obtained using the image-guided system to assess the coefficient of variation, the within-subject standard deviation and intraclass correlation coefficient (ICC). A paired Student t test was used to assess statistical significance between the Verion and the auto-refracto-keratometer. A Pearson’s correlation coefficient (r) was obtained for all measurements, and the level of agreement was verified using Bland–Altman plots. Results: The right eyes of 73 patients were evaluated by each platform. The Verion coefficient of variation was 0.3% for the flat and steep keratometry, with the ICC being greater than 0.9 for all parameters measured. Paired t test showed statistically significant differences between groups (P = 0.0001). A good correlation was evidenced for keratometry values between platforms (r = 0.903, P = 0.0001 for K1, and r = 0.890, P = 0.0001). Bland–Altman plots showed a wide data spread for all variables. Conclusion: The image-guided system provided highly repeatable corneal power and keratometry measurements. However, significant differences were evidenced between the two platforms, and although values were highly correlated, they showed a wide data spread for all analysed variables; therefore, their interchangeable use for biometry assessment is not advisable.

Publication Information

Output type

Research Output:
Contribution to journal
Article
Peer-review

Original language

English

Pages from-to (Number of pages)

Pages 951-957 (7 pages)

Journal (Volume, Issue Number)

International Ophthalmology (Volume 38, Issue 3)

Publication milestones

  • Published - 01/06/2018

Publication status

Published - 01/06/2018

ISSN

0165-5701

Publication IDs

  • Scopus: 85018699518
  • WOS: 000434373500007