Comparison of biometric measurements obtained by the Verion Image-Guided System versus the auto-refracto-keratometer
- Cecilio Velasco-Barona,
- Guadalupe Cervantes-Coste,
- Erick Mendoza-Schuster,
- Claudia Corredor-Ortega,
- Nadia L. Casillas-Chavarín,
- Alejandro Silva-Moreno
- Asociacion para Evitar la Ceguera en Mexico
Publication metrics
Metrics
Abstract
Purpose: To compare the biometric measurements obtained from the Verion Image-Guided System to those obtained by auto-refracto-keratometer in normal eyes. Methods: This is a prospective, observational, comparative study conducted at the Asociación para Evitar la Ceguera en México I.A.P., Mexico. Three sets of keratometry measurements were obtained using the image-guided system to assess the coefficient of variation, the within-subject standard deviation and intraclass correlation coefficient (ICC). A paired Student t test was used to assess statistical significance between the Verion and the auto-refracto-keratometer. A Pearson’s correlation coefficient (r) was obtained for all measurements, and the level of agreement was verified using Bland–Altman plots. Results: The right eyes of 73 patients were evaluated by each platform. The Verion coefficient of variation was 0.3% for the flat and steep keratometry, with the ICC being greater than 0.9 for all parameters measured. Paired t test showed statistically significant differences between groups (P = 0.0001). A good correlation was evidenced for keratometry values between platforms (r = 0.903, P = 0.0001 for K1, and r = 0.890, P = 0.0001). Bland–Altman plots showed a wide data spread for all variables. Conclusion: The image-guided system provided highly repeatable corneal power and keratometry measurements. However, significant differences were evidenced between the two platforms, and although values were highly correlated, they showed a wide data spread for all analysed variables; therefore, their interchangeable use for biometry assessment is not advisable.
Publication Information
Output type
Original language
EnglishPages from-to (Number of pages)
Pages 951-957 (7 pages)Journal (Volume, Issue Number)
International Ophthalmology (Volume 38, Issue 3)Publication milestones
- Published - 01/06/2018
Publication status
ISSN
0165-5701Publication IDs
- Scopus: 85018699518
- WOS: 000434373500007