Comparative Study of Two Infrared Meibographers in Evaporative Dry Eye Versus Nondry Eye Patients
- Fernando Garduño,
- Anna Salinas,
- Katia Contreras,
- Yadira Rios,
- Naomi García,
- Patricio Quintanilla
- Universidad de Monterrey,
Research Output:
Contribution to journal
Article
Peer-reviewPublication metrics
Metrics
SciVal
FWCI
0.65
SciVal
Author count
8
SciVal
Paper percentile
59
SciVal
Citations
5
PlumX, opens in new tab
Captures
8
Citations
5
Abstract
Objective: To compare the measurement of the meibomian gland (MG) dropout between two infrared meibographers in patients with and without dry eye.
Methods: The right eyelids of each patient were imaged using the Antares and Cobra meibography devices. All images were analyzed using Phoenix software to calculate the percentage of the MG dropout. Lipid layer thickness, eyelid margin characteristics, ocular surface staining, MG secretion, number of expressible glands, and noninvasive tear breakup time were also evaluated. A comparison between nondry eye and evaporative dry eye was performed.
Results: Eighty participants (mean age, 36.93 years and 51.3% women) were included, of which 67.5% had nondry eye. A significant difference was observed in the dropout percentage of the superior eyelid between the Antares and Cobra devices (P=0.007) for all participants and when only the nondry eye patients were examined. In patients with dry eye, no significant differences were found in the dropout measurements of both eyelids.
Conclusions: Statistically significant differences in the MG dropout percentage in the upper eyelid of nondry eye patients were obtained from both meibographers. The measurements were similar in patients with dry eye, suggesting that the two instruments can be interchanged.
Methods: The right eyelids of each patient were imaged using the Antares and Cobra meibography devices. All images were analyzed using Phoenix software to calculate the percentage of the MG dropout. Lipid layer thickness, eyelid margin characteristics, ocular surface staining, MG secretion, number of expressible glands, and noninvasive tear breakup time were also evaluated. A comparison between nondry eye and evaporative dry eye was performed.
Results: Eighty participants (mean age, 36.93 years and 51.3% women) were included, of which 67.5% had nondry eye. A significant difference was observed in the dropout percentage of the superior eyelid between the Antares and Cobra devices (P=0.007) for all participants and when only the nondry eye patients were examined. In patients with dry eye, no significant differences were found in the dropout measurements of both eyelids.
Conclusions: Statistically significant differences in the MG dropout percentage in the upper eyelid of nondry eye patients were obtained from both meibographers. The measurements were similar in patients with dry eye, suggesting that the two instruments can be interchanged.
Publication Information
Output type
Research Output:
Contribution to journal
Article
Peer-reviewOriginal language
EnglishPages from-to (Number of pages)
Pages 335-340 (6 pages)Journal (Volume, Issue Number)
Eye and Contact Lens (Volume 47, Issue 6)Publication milestones
- E-pub ahead of print - 24/11/2020
- Published - 01/06/2021
Publication status
Published - 01/06/2021
ISSN
1542-2321Publication IDs
- Scopus: 85106667629
- PubMed: 33239503