Comparison of the measurement of central corneal thickness measured by a new device of tomography with Scheimpflug camera and Placido rings (Sirius ®) and ultrasonic pachymetry in healthy subjects

Original title: Comparación de la medición del grosor corneal central medido con un nuevo equipo de tomografía con cámara de Scheimpflug y anillos de Plácido (Sirius ®) y paquimetría ultrasónica en sujetos sanos
  • Fundación Destellos de Luz
    ,
  • Instituto Mexicano del Seguro Social
    ,
  • Instituto Nacional de Salud Publica
Research Output:
Contribution to journal
Article
Peer-review

Open access

Publication metrics

Metrics

SciVal
Author count
3
SciVal
Paper percentile
31

Abstract

Objective: To assess the repeatability and comparability of the central corneal thickness measurements obtained from a novel Scheimpflug camera with a Placido-disc topographer (Sirius ®) and ultrasound pachymetry. Material and methods: In a sample of thirty unoperated eyes of healthy subjects, the central corneal thickness was analyzed by a single examiner using the Sirius ® and the ultrasound pachymeter. Repeatability of three sets of measurements from each device were evaluated using the coefficient of variation, standard deviation, and intraclass correlation coefficient. T-test with Bonferroni adjustment and Bland and Altman plots were used to assess agreement between devices. Results: For the Sirius ® the coefficient of variation was 0.291%, and for the ultrasound pachymeter was 0.238%. The standard deviation was 0.123μm and 0.101μm for Sirius ® and US respectively and the intraclass correlation coefficient was higher than 0.95 in all measurements. Ultrasound pachymetry systematically yielded higher values and statistically significant (mean 543.867 vs 558.267μm Sirius ® and ultrasound pachymeter respectively), but the difference was not clinically significant. Conclusion: In the assessment of normal corneas, central corneal thickness measured with both devices showed high repeatability and good agreement. Even though central corneal thickness measurements obtained by Sirius ® were systematically lower than those obtained by ultrasound pachymeter, they do not have clinical significance.

Publication Information

Output type

Research Output:
Contribution to journal
Article
Peer-review

Original language

Spanish

Pages from-to (Number of pages)

Pages 225-229 (5 pages)

Journal (Volume, Issue Number)

Revista Mexicana de Oftalmologia (Volume 89, Issue 4)

Publication milestones

  • Published - 10/2015

Publication status

Published - 10/2015

ISSN

0187-4519

Publication IDs

  • Scopus: 84945960589
  • ORCID: /0000-0001-9515-1992/work/21895505