Comparación de la medición del grosor corneal central medido con un nuevo equipo de tomografía con cámara de Scheimpflug y anillos de Plácido (Sirius ®) y paquimetría ultrasónica en sujetos sanos

Manuel Garza-León, Paola de la Parra-Colín, Tonatiuh Barrientos-Gutierrez

Resultado de la investigaciónrevisión exhaustiva

Resumen

Objective: To assess the repeatability and comparability of the central corneal thickness measurements obtained from a novel Scheimpflug camera with a Placido-disc topographer (Sirius ®) and ultrasound pachymetry. Material and methods: In a sample of thirty unoperated eyes of healthy subjects, the central corneal thickness was analyzed by a single examiner using the Sirius ® and the ultrasound pachymeter. Repeatability of three sets of measurements from each device were evaluated using the coefficient of variation, standard deviation, and intraclass correlation coefficient. T-test with Bonferroni adjustment and Bland and Altman plots were used to assess agreement between devices. Results: For the Sirius ® the coefficient of variation was 0.291%, and for the ultrasound pachymeter was 0.238%. The standard deviation was 0.123μm and 0.101μm for Sirius ® and US respectively and the intraclass correlation coefficient was higher than 0.95 in all measurements. Ultrasound pachymetry systematically yielded higher values and statistically significant (mean 543.867 vs 558.267μm Sirius ® and ultrasound pachymeter respectively), but the difference was not clinically significant. Conclusion: In the assessment of normal corneas, central corneal thickness measured with both devices showed high repeatability and good agreement. Even though central corneal thickness measurements obtained by Sirius ® were systematically lower than those obtained by ultrasound pachymeter, they do not have clinical significance.

Título traducido de la contribuciónComparison of the measurement of central corneal thickness measured by a new device of tomography with Scheimpflug camera and Placido rings (Sirius ®) and ultrasonic pachymetry in healthy subjects
Idioma originalSpanish
Páginas (desde-hasta)225-229
Número de páginas5
PublicaciónRevista Mexicana de Oftalmologia
Volumen89
N.º4
DOI
EstadoPublished - oct 2015

All Science Journal Classification (ASJC) codes

  • Ophthalmology

Huella Profundice en los temas de investigación de 'Comparación de la medición del grosor corneal central medido con un nuevo equipo de tomografía con cámara de Scheimpflug y anillos de Plácido (Sirius <sup>®</sup>) y paquimetría ultrasónica en sujetos sanos'. En conjunto forman una huella única.

Citar esto